
-86°C hydrocarbon ULT freezer
- Temperature Control System : Controlled by a micro-computer, LCD display, touch screen eye level control panel; The inner temperature can be adjusted at the range of -40°C~-86°C, and the temperature regulation accuracy is 0.1°C. ambient temp +15 to +32°C.
- Security Control System : Multi malfunction alarms: high-temperature alarm, low-temperature alarm, sensor malfunction alarm, condenser dirty alarm, low battery alarm, power failure alar, overpressure alarms, over ambient temperature alarms, open the door alarms and so on for safe use; Besides, you can check history alarming information and derive data. All the history data can be saved over 10 years; Three alarming methods: buzzer alarming, independent icon alarming, and remote alarming; Protection function: compressor power-on
delay protection, time between power off; overpressure protection, password protection and normal operation protection under failure mode, door lock. Password protection can prevent parameters from being changed randomly. Display screen with running indication and communication status indication can show real-time operation of the freezer; The power switch has two-way fuse over-current protection, together with the power-on indicator light ensure safety.
Refrigeration System : Optimized cascade refrigeration technology, special evaporation system and condensation system, strong refrigerating efficiency, CFC HCFC free natural Refrigeration International famous brand compressor and EBM fan electromotor and oil separator, stable product performance, strong refrigerating efficiency; Enhanced sealing design: double door seal.
Humanized Design : Unique handle design: the door can be opened by both hands, individual press lock on the handle and 2 padlocks; Wide voltage design can help it work stably under 198-254V; Can show the temperature curve and history data, and have derive data function; Press-out type dust removal screen with locks, simple and convenient for maintenance; Thermometer holes design for temperature examination, Battery backup for control system up to 24 hours at to power failure.
Individual Removable Foam Inner Doors
- Individual inner doors can be opened independently to minimize frost buildup inside the chamber.
- Unique door seal design for the minimum loss of cold temperature during a door opening.
- Compatibility with existing racking systems from competitors.
- Stainless steel handle to ensure proper strength for door latching.
- Some interior door handles have been upgraded to stainless steel for a more comfortable door-opening experience.
Excellent Door Seals
Total of five gaskets to safeguard the freezer temperature, including four seals for the exterior door and one for each inner door.
Touch Screen
- 7″ LCD display for simple operation and visible running status.
- Optimized slideway design for easier door opening and closing.
- Two port holes for ease of temperature monitoring.
High-Efficiency Refrigeration Components
- Hermetically sealed industrial-grade compressors for low sound levels.
- High-efficiency oil separator.
- Unique hydrocarbon refrigeration system to provide maximum ultra-low refrigeration application efficiency in extreme conditions.
Energy Saving Refrigeration
High-efficiency cooling fans and compressors, combined with hydrocarbon refrigerants, ensure energy savings and long-term sample security.
Improved Handle Design
Lockable handle with a unique key prevents other Thermolab freezer owners’ access to your precious samples. Also comes with space for a padlock for extra security.
Pressure Equalization Port
Heated port with a spring-assisted mechanism to prevent icing on the vent allows users to reopen the main door soon after entering. Adopts chromium plating, rust-proof.
USB Interface
Enables users to download historical temperature data for compliance/auditing purposes.
Multilayered Sealing Structure
Triple layer of gaskets split between main and inner doors decreases heat loss and guarantees excellent warm-up times in the event of a power failure.
Multilevel Alarms
Alarming functions that include high and low temperature, sensor error, power failure, high ambient center, and door ajar.
Dual Cascading
Dual cascade refrigeration system with height adjustable shelves – TW-86L450T / TW-86L570T / TW-86L710T.
Model | TW-86L450T | TW-86L570T | TW-86L710T | Old Model TW-86L725T | |
---|---|---|---|---|---|
Technical Data | |||||
Cabinet Type | Upright | ||||
Climate Class | SN | ||||
Cooling Type | Direct Cooling | ||||
Defrost Mode | Manual | ||||
Refrigerant | HC (R290/R1700) | ||||
Sound Level (dB(A)) | 50 dB | ||||
Performance | |||||
Cooling Performance | -86°C | ||||
Temp Range | -40°C to -86°C | ||||
Control | |||||
Controller | Microprocessor / PLC 7 inches Touch screen HMI | ||||
Display | LCD | ||||
Electrical Data | |||||
Power Supply (V/Hz) | 230V / 50Hz | ||||
Electrical Current (A) | 7 | ||||
Power (W) | 1200 | 1200 | 1200 | ||
Construction | |||||
Capacity (L/Cu. Ft.) | 450/16 | 570/20.2 | 725/25.6 | ||
Made of Construction | Internal – SS 304 / PPGI (as per user requirement) External – SS 304 / CRCA Powder Coated (as per user requirement) |
||||
Net / Gross Weight (kg) | 300/330 | 317/395 | 359/395 | ||
Alarms | |||||
Remote Alarm (Dry Contacts) | Yes | ||||
High/Low Temp | Yes | ||||
Hot Condenser | Yes | ||||
Power Failure | Yes | ||||
Sensor Error | Yes | ||||
Low Battery | Yes | ||||
High Ambient Temp. | Yes | ||||
Door Ajar | Yes | ||||
Certification | |||||
ISO 9001, ISO 13485, IAF Body, NABL, European CE with IEC test std ISO 60601 / 610101, WHO-GMP & ISO 14001, ISO 40001, ROHS. |